6.3 “Error” on the Display
An “Error” message is displayed on the KT-10 v2’s screen when it fails to compute a
result as seen in the figure 128.
Figure 128: Measurement Error
The main reason for this error message may be that a measurement sequence is not being
followed properly.
To ensure that a correct measurement sequence is being followed, please try do the
following:
1.
Set the KT-10 v2 in the Measure mode.
2.
With the KT-10 v2 at arm's length, in the free air, press the UP button.
Meter beeps, and the "progress bar" starts to run.
3.
Immediately after the first beep, quickly move the KT-10 v2 to the
sample’s surface and press the UP button again.
4.
Hold the KT-10 v2 on the sample until it beeps a second time, with a
different tone. After the second beep, quickly move the instrument back
to arm's length. Do not push the button again.
5.
The KT-10 v2 beeps a third time on it's own; displaying result on the
screen.
TIP:
Alternately, the Scanner mode can be used to confirm the operation of the meter.