SYMBOL
CAUSE
CORRECTION
persists, contact the customer
service.
Er 6
The sensor is sensing
great fluctuation in the
pressure
Remain calm during the
measurement and do not move.
Er 7
The pressure that the
sensor sensing is over
the limit
If it happens occasionally, repeat
the measurement; if this always
happens, contact the customer
service.
Er 8
Electronic failure
Please contact customer service.
5.9 Troubleshooting
Problem
Check
Cause and solution
No power, No
display appears
Check the batteries status
Replace with new ones
Check the polarity positon
Insert the batteries with
the correct placement.
No inflation
Check the correct insertion
of the tube connector
Insert into the air socket
tightly
Check that cuff or tube
hasn't air leak
Replace cuff
Check if the plug is
brocken or it has a leak
Replace cuff
Error display and
stop working
Arm has been moved
during inflation
Keep the body peaceful
Chack if chatting when
measure is in process
Keep quite when
measure
Cuff leak
Check if the cuff wrap is
too loose
Wrap the cuff tightly
Check if the cuff is broken
Replace cuff
Cuff does not
tighten
Check the cuff has fasten
too loose
Fasten cuff more tight
Check the cuff isn't
brocken
Replace cuff
No deflate
Display may show an error
code
Open the cuff, switch
appliance OFF, wait 10
minutes and test again; if
error is repeated, pls
contact service center
Appliance switch on
but doesn't work
Display may show an error
code
Check error codes at
previous table and/or
contact customer service.
Applince switches
off during
measurement
Battery almost empty
Replace with new ones
52
Содержание 8028153111291
Страница 1: ...MISURATORE DI PRESSIONE BLOOD PRESSURE MONITOR Libretto Istruzioni Instruction Manual 1 ...
Страница 12: ...3 COMPONENTI DEL VOSTRO MISURATORE PRESSIONE 3 1 Unità principale 12 ...
Страница 29: ...29 ...
Страница 30: ...30 ...
Страница 31: ...31 ...
Страница 42: ...3 COMPONENTS OF YOUR BLOOD PRESSURE MONITOR 3 1 Measuring unit 42 ...
Страница 57: ...57 ...
Страница 58: ...58 ...