Chapter 12 - System Safety and Maintenance
Acoustic Output Indices
Terason t3000 / Echo Ultrasound System User Guide
275
The
relative error in the spatial-peak derated temporal average intensity for
unscanned modes
, defined as
SPTA.3-unsc.
, is approximately:
where the following is defined:
The
relative error in estimating spatial-peak derated temporal average intensity for
scanned modes
, defined as
SPTA.3-scan.
, is approximately:
where the following are defined:
The
relative error in the peak derated rarefactional pressure
, defined as
pr.3
, is
approximately:
The
relative error in the Mechanical Index
, defined as
MI
, is approximately:
lin
Relative error due to the effects of non-linearities in the
pressure waveform upon the hydrophone and integral
amplifier, estimated as 0% for this transducer set
PD
Relative error in estimate of pulse duration, estimated as 5%
prf
Relative error in prf estimation for operating condition giving
actual peak ISPTA.3-unsc., estimated as 1%
srf
Relative error in estimating srf (scan repetition frequency) for
operating condition giving actual peak I
SPTA.3-scan.
,
estimated as 1%
BOF
Relative error in estimating BOF (beam-overlap factor) for
operating conditions giving actual peak I
SPTA.3-scan.
,
estimated as 5%
Note:
In accordance with Subsection 6.4.3 (Measurement of Precision of Peak
Rarefactional Pressure, p
r
) of the
Standard for Real-Time Display of Thermal and
Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
, a series
of ten independent measurements on the specified standard test transducer/driver
combination yielded a relative standard deviation of W of less than 1% for the
sample measurements.
SPTA.3-unsc.
2
SPPII.3
2
prf
1 2
/
11.3%
=
+
=
SPTA.3-scan.
2
SPII.3
2
srf
2
BOF
+
+
1 2
/
12.8%
=
=
pr.3
1
2
---
SPPII.3
5.9%
=
=