Overload protection: 250V DC or AC rms at all ranges.
At 20M
range, short circuit test lead, LCD displaying
12 digits is normal. During measurement minus these
12 digits from the obtained reading.
When measuring 20
and 200 range, short circuit test
clips to display the resistance value of the test lead.
Subtract this value from the measurement value to obtain
the correct tested value.
l
l
l
Accuracy: (a% r b digits),guarantee for 1 year.
Operating temperature: 23
o
C 5
o
C.
Relative humidity: < 75%.
Temperature coefficient: 0.1 x (specified accuracy) / 1
o
C
Model 72-8155: OPERATING MANUAL
20
Accuracy Specifications
A. Resistance Test
Range Resolution
Accuracy
200
2k
20k
200k
2M
20M
0.1
1
10
100
1k
10k
Remarks:
(0.8%+3)
(0.8%+1)
[2%(rdg-12)+5]