To test the inductance, please follow the following procedure:
Set the rotary switch to
Lx
measurement mode.
If the tested inductance value is unknown, use the
maximum measurement position and decrease the range
step by step until a satisfactory reading is obtained.
Insert the test clips into the corresponding
Lx
input
terminals.
The measured value shows on the display.
When measuring inductors in the 2mH range, you should
first short circuit the test leads and note the measured
inductance value of the leads. Then the
correct reading is the measured reading, minus the short
circuit reading.
The small value jack on the surface of the meter should
be used when measuring small value inductors.
The Meter cannot check the quality of the inductance.
When inductance measurement has been completed,
disconnect the connection between the test clips and
the circuit under test and remove the test clips from the
input terminals of the Meter.
l
l
l
l
Model 72-8155: OPERATING MANUAL
17
D. Inductance Measurement
(see figure 6)
(figure 6)
1.
2.
3.
4.
Remarks:
Содержание 72-8155
Страница 2: ...Model 72 8155 OPERATING MANUAL 2 ...