Function Menu
Settings
Description
Reject high-frequency
components of the triggering
signal (over 80kHz signals)
High-frequency
rejection
Reject low-frequency
components of the triggering
signal (below 80kHz signals)
Trigger coupling (F5)
Low-frequency
rejection
Set trigger on the signal
rising edge
Set trigger on the signal
falling edge
Set trigger on the signal both
rising and falling edge
Rising
Falling
Rise/fall
Slope (F3)
Set to automatic trigger. The
DSO will continuously
perform data acquisition
without triggering signal.
Set to normal trigger. The
DSO will only perform data
acquisition when there is a
triggering signal.
Set to single trigger. The
DSO will only perform 1
cycle of the data acquisition
when there is a triggering
signal.
Mode (F4)
Automation
Normal
Single
Block the DC component of
the triggering signal
Pass through DC and AC
components of the triggering signal
DC
AC
Trigger
coupling (F5)
Pulse width trigger
For pulse width trigger, the trigger time shall be subject
to the pulse width of the triggering signal. You can capture
abnormal pulse by setting pulse width conditions.
Table 2-8
Type
Pulse width
Set CH1 as the triggering
signal source
Set CH2 as the triggering
signal source
Set external trigger input
channel as the triggering
signal source
Source (F2)
EXT
CH2
CH1
26
72-2650/72-10510/72-8225A /72-8705A/72-8710A user manual
Содержание 72-10510
Страница 1: ...72 2650 72 8705A 72 8710A 72 8225A 72 10510 P N 110401105556X REV 0 DATE 2016 02 20 ...
Страница 3: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 14: ...DSO User Interface Menu Display 72 10510 pictured 4 72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 73: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...