Pass through DC and AC
components of input signal.
Disconnect input signal.
DC
Grounding
Coupling
Bandwidth
Limited
Limit bandwidth to 20MHz to reduce
displayed noise. Full bandwidth.
Turn off bandwidth limited.
On
Off
On
Off
V/div
Rough
adjustment
Fine
adjustment
Set vertical deflection factor by
coarse adjustment based on 1-2-5
scale. Rough adjustment setting
range is further subdivided for fine
adjustments to improve vertical
resolution.
Probe
1×
10×
100×
1000×
Please set the probe compensation
system ratio to match the probe
attenuation coefficient of the probe
to be used.
Inverting
Activate reverse function of
waveform. Waveform will be
displayed reversely.
Waveform is displayed normally.
Setting coupling channel
The following example shows how to change the coupling
of the input channel. If the measured signal connected to the
CH1 channel is a sinusoidal signal containing DC component,
press F1 to select AC coupling, to set the CH1 into AC
coupling mode. The DC component of the measured
signal is blocked.
The Waveform display is shown in the figure below.
If you press F1 to select DC coupling, setting both
the DC and AC components of measured signals at
Figure 2-1 Blocked DC Component of Signal
13
72-2650/72-10510/72-8225A /72-8705A/72-8710A user manual
Содержание 72-10510
Страница 1: ...72 2650 72 8705A 72 8710A 72 8225A 72 10510 P N 110401105556X REV 0 DATE 2016 02 20 ...
Страница 3: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 14: ...DSO User Interface Menu Display 72 10510 pictured 4 72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 73: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...