of the trigger threshold on the input signal edge.
Pulse Width trigger:
trigger produces when the pulse
width of the trigger signal satisfies some set trigger >, <
or = condition.
Video Trigger:
conduct field or line trigger for standard
video signal. The triggering can be enabled by pressing
the [F1] button at the [TRIG MENU] for the 72-10510/
72-8225A/72-8705A/72-8710A DSO, by pressing [F1]
button at the [MENU] for the 72-2650 DSO;
Alternating trigger:
applicable to trigger signals without
frequency correlation.
All kinds of trigger menus shall be respectively
described as follows.
The user can enable the alternating trigger by pressing [F2]
button at the [TRIG MENU] for the 72-10510 /72-8225A/
72-8705A/72-8710A DSO, or by pressing [F2] button at
the [MENU] for the 72-2650 DSO.
Edge trigger
Edge trigger mode refers to the triggering of trigger
threshold on the input signal edge. When selecting “edge
trigger”, trigger on the rising edge or falling edge of input
signal shall be completed.
Function Menu
Settings
Description
Type
Table 2-7
Edge
CH1
CH2
Set CH1 as the triggering
signal source
Set CH2 as the triggering
signal source
Set external trigger input
channel as the triggering
signal source
EXT
Set external trigger source
(divided by 5) as trigger level
range (Only for 72-2650)
EXT/5
Set AC power line as
trigger source
AC Line
Set to Alternative trigger
between CH1 and CH2 as
signal source
Alter
Source (F2)
25
72-2650/72-10510/72-8225A /72-8705A/72-8710A user manual
Содержание 72-10510
Страница 1: ...72 2650 72 8705A 72 8710A 72 8225A 72 10510 P N 110401105556X REV 0 DATE 2016 02 20 ...
Страница 3: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 14: ...DSO User Interface Menu Display 72 10510 pictured 4 72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 73: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...