72-10510/72-8225A/72-8705A/72-8710A
Figure 2-17 Function Key of Sampling System
While ACQUIRE button is pressed, the function menu
will change to set the data acquisition setting for the DSO.
Table 2-11 Acquire Menu
Function
Menu
Settings Description
Set to sampling acquisition mode.
Set to peak value acquisition mode.
Set to average acquisition mode.
Average
Peak value
Sampling
Acquisition
Mode (F1)
At the average sampling mode, set
average number by using the
multipurpose knob (Average number
can be set to 2,4,8,16,32,64,128,256)
2~256
Averages
(@)
Set to Real-time sampling.
Set to Equivalent sampling.
Real-time
Equivalence
Sampling
(
F3
)
Turn on fast acquiring. Waveform
capturing rage up to at 2k
waveform/second.
Turn off fast acquiring.
ON
OFF
Fast Acq
(
F4
)
By changing the acquisition setting of the DSO, the user
can observe the input signal in a different manner. The
example below is a noise waveform acquired by different
acquisition setting of the 72-2650/72-10510/72-8225A /
72-8705A/72-8710A DSO.
Figure 2-18 shows the original signal using sampling
acquisition, the user can see that the waveform display is
very noisy.
Figure 2-19 shows the same signal with average
acquisition of a 32 number, please observe that the
waveform display becomes much smoother.
32
72-2650/72-10510/72-8225A /72-8705A/72-8710A user manual
Содержание 72-10510
Страница 1: ...72 2650 72 8705A 72 8710A 72 8225A 72 10510 P N 110401105556X REV 0 DATE 2016 02 20 ...
Страница 3: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 14: ...DSO User Interface Menu Display 72 10510 pictured 4 72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...
Страница 73: ...72 2650 72 10510 72 8225A 72 8705A 72 8710A user manual ...