How to Use EDGEVue and EDGECloud
42
Watch Mode settings
Manual start
: enables you to activate
Watch Mode from the
Scan
tab. With
this option selected, an ON/OFF
button is displayed on the
Scan
screen.
Slide the button to the right to activate Watch Mode (
).
Disabled
: Watch Mode is disabled by default. Tap on
Manual start
to
activate this function.
With Watch Mode activated, the
latest temperature reading is
updated once per minute and
displayed in the list on the
Scan
tab.
When Watch Mode is activated, the colored icons on the Scan tab have a
slightly different meaning than for “Discover mode.”
Green indicates a programmed module, with datalogging
started or not, whose last-read value has not exceeded its limit
value (that is, currently not in an alarm state).
Red indicates a programmed module, currently in an alarm
state, whose last-read value has exceeded its limit value. If a
delay was configured, it has been reached. EDGEVue plays an
alarm sound once a minute, for several seconds, if the latest
reading on one or more modules has exceeded an alarm limit.
Grey indicates either a non-programmed module or one on
which there was an error reading the sensor (i.e. the current
reading is not available). The text beneath the module name
shows the status (“Not started” or “Last reading: N/A”).