How to Use EDGEVue and EDGECloud
31
Setting high and low limits
You may set high and low limits in the
Alarms
tab to ensure that the
temperature in the monitored
environment remains within a specific
range.
The default values in the fields show
the module’s available temperature
range. Tap on checkboxes to enable
options and in fields to adjust settings.
Low limit
: the lowest acceptable
temperature reading.
High limit
: the highest acceptable
temperature reading.
Low delay
: how long the reading is
lower than the Low limit (multiple of
the reading interval, in minutes only).
High delay
: How long the reading is
higher than the High limit before
triggering an alarm (multiple of the
reading interval, in minutes only).
If an alarm occurs at any point after the start of the mission, the module
icon changes color to show an alarm in the
Scan
list (unless Watch Mode is
activated, as described in
Monitoring with Watch Mode,
page
41
).
High and low temperature
range limits
Alarm condition
indicated by red icon