
May
2015
WorkHorse H-ADCP Operation Manual
Page 22
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
H-ADCP Orientation
The H-ADCP must be mounted with its transducer orientation in the horizontal direction. That is, the
three transducers, which are for water velocity measurement, must be looking horizontally (also called
side-looking). The pressure sensor, which is for water depth measurement, must be looking up towards
the water surface.
X-Axis - Pitch is measured around the X-axis and should be set to zero.
Y-Axis - Roll is measured around the Y- axis and should be set to zero.
Pressure Sensor – The pressure sensor must be facing up towards the water surface (for the 300 kHz
NB H-ADCP, it is facing the beams – Beam 3 should be upward).
Figure 11.
Installing the H-ADCP
I/O CABLE
48 VDC
H-ADCP
SurfaceView
WavesMon