WorkHorse H-ADCP Operation Manual
May
2015
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 143
Table 18:
Polled Mode Commands
Command
Description
!
Execute a Break reset
+
Increment internal clock by 1 second
-
Decrement internal clock by 1 second
D
Dump the last ensemble
E
Print the current ensemble number
T
Print the current time
The commands are not echoed and they don’t need to be followed by a CR/LF pair. These
commands are only available when CP = 1.
Enabling polled mode disables the battery saver feature. Do not enable this mode when running
from batteries.
CQ – Transmit Power
Purpose
Allows the transmit power to be adjusted.
Format
CQnnn
Range
nnn = 0, or 1 to 255 (0 = Low, 1 to 255 = High)
Default
CQ255
Recommended Setting.
The default setting for this command is recommended for most applications.
Description
Allows the transmit power to be set high or low.
This command only affects 75 kHz Long Ranger systems.
Although other frequency H-ADCPs allow the CQ command to be set to values other than the default,
the CQ command has no effect and is not supported by
PlanADCP
.