WaveRunner Xi-A and MXi-A Oscilloscopes
90
922138-00 Rev A
Aux Output
In addition to a calibration signal, the following signals can be output through
the
AUX OUTPUT
connector:
Square Wave
Trigger Out
- can be used to trigger another
oscilloscope
DC level
- a reference level
Trigger Enabled
- can be used as a gating
function to trigger another instrument when the
oscilloscope is ready
Pass/Fail
- allows you to set a pulse duration
from 1 ms to 500 ms. This pulse occurs when
pass/fail testing is active and conditions are met.
Aux Output Off
- turns off the auxiliary output
signal
Auxiliary Output Setup
Touch
Utilities
Utilities Setup...
from the menu bar.
Touch the
Aux Output
tab.
If you simply want a 1 kHz, 1 V square wave, touch the button so labeled.
If you want a specialized output, touch one of the buttons under
Use Auxiliary
Output For
.
Touch inside the
Amplitude
field and enter a value, using the pop-up numeric
keypad. If you want a TTL level signal, touch the
TTL Level
checkbox. The
Amplitude
field becomes unavailable.
If you selected Square Wave, touch inside the
Frequency
field and enter a
value, using the pop-up keypad. You can set a value from 5.0 Hz to 5 MHz.
If you selected Pass/Fail, touch inside the
Pulse Duration
field and enter a
value from 1 ms to 500 ms, using the pop-up numeric keypad.
Содержание waverunner mxi-a series
Страница 1: ...Getting Started Manual WaveRunner Xi A WaveRunner MXi A Series Oscilloscopes...
Страница 2: ......
Страница 3: ...WaveRunner Xi A MXi A Series Oscilloscopes Getting Started Manual January 2013...
Страница 22: ...WaveRunner Xi A and MXi A Oscilloscopes 14 922138 00 Rev A Front Panel Controls...
Страница 126: ...WaveRunner Xi A and MXi A Oscilloscopes 118 922138 00 Rev A...
Страница 127: ......
Страница 128: ......