Getting Started Manual
922138-00 Rev A
25
B
At this point, the following dialog is shown at the bottom of the display, and
the grid area shortens until this dialog is closed (by touching the
Close
button
on the upper right hand corner of this dialog).
Coupling
Coupling choices are as follows:
DC 50 Ω
GROUND
DC 1 MΩ
AC 1 MΩ
Select coupling by touching inside the
Coupling
field and choosing a coupling
mode from the pop-up menu.
NOTE:
The coupling choices for a channel change if a ProBus probe is
connected to a channel.
Deskew
Deskew
allows you to compensate for different lengths of cables, probes, or
anything else that might cause timing mismatches between signals. Connect
all probes to the desired channels, then probe a common signal with each
probe and adjust for timing differences using
Deskew
.
Probe Attenuation
If you use a Teledyne LeCroy ProBus compatible active probe, or a probe
compatible with Probe Ring, the attenuation is automatically set by the
oscilloscope. If it is not automatically set, select a value here by touching
inside the
Probe Atten
field selecting a value from the pop-up menu.