B
Custom Measurements with My Measure
You can choose to customize up to six parameters by touching
My Measure
and then selecting the measurements desired.
Status Symbols
Below each parameter appears a symbol that indicates the status of the
parameter, as follows:
A green check mark means that the oscilloscope is returning a
valid value.
A crossed-out pulse means the oscilloscope is unable to determine
top and base. However, the measurement could still be valid.
A downward pointing arrow indicates an underflow condition.
An upward pointing arrow indicates an overflow condition.
An upward-and-downward pointing arrow indicates an underflow
and overflow condition.
Statistics
You can turn on statistics by checking the
Statistics On
checkbox in the Measure main
dialog. A table of values is then displayed below
the grid.