Getting Started Manual
922138-00 Rev A
79
B
Touch the Summary View to enable a line of text that shows concisely the
status of your last waveform and keeps a running count of how many sweeps
have passed.
Touch inside the
Pass If
field, and select a Boolean condition from the pop-up
menu.
If you want to set up the test to end after a finite number of sweeps, touch the
Stop Test
checkbox. Then touch inside the
After
field and enter a value, using
the pop-up numeric keypad.
Under
If
, touch either the
Pass
or
Fail
button to set the actions to
occur upon your waveform's passing or failing the test.
Under
Then
, touch the actions you want to occur: stop test, sound
alarm, print result, emit pulse, or save the waveform. If you want to
have the results printed and your oscilloscope is not equipped with a
printer, be sure that the it is connected to a local or network printer. See
Printing.
If you want to save your waveform automatically, touch the
Save Setup
. This
takes you out of the current dialog and opens the
Save Waveform
dialog.
Refer to the
Saving and Recalling Waveforms
section of this manual for more
details.
Test your Pass/Fail conditions by touching the
Force Actions Once
button.
Press the
Clear All
button to quickly uncheck all checkboxes if you want to
change your selections.
Содержание waverunner mxi-a series
Страница 1: ...Getting Started Manual WaveRunner Xi A WaveRunner MXi A Series Oscilloscopes...
Страница 2: ......
Страница 3: ...WaveRunner Xi A MXi A Series Oscilloscopes Getting Started Manual January 2013...
Страница 22: ...WaveRunner Xi A and MXi A Oscilloscopes 14 922138 00 Rev A Front Panel Controls...
Страница 126: ...WaveRunner Xi A and MXi A Oscilloscopes 118 922138 00 Rev A...
Страница 127: ......
Страница 128: ......