Sequence Mode Display
Set up a Sequence Mode display by first enabling
Sequence
mode in the
Timebase Horizontal dialog. A
Num Segments
value must also be provided as
follows:
Touch the Timebase descriptor label.
In the
Timebase
dialog, touch the
Sequence
mode
button.
Select the
Sequence
tab. Under
Acquisition Settings
,
touch inside the
Num Segments
field and enter a
value from 2 to 500.
Under Touch inside the
Timeout
field and enter a
value from 10 ms to 100 s. Then check the
Enable
Timeout
checkbox.
Under Display Settings, touch inside the
Display
Mode
field and select a mode from the pop-up
menu.
NOTE:
The maximum value that you can enter for
Starting at
depends on the
Num Segments
value you entered under
Acquisition Settings
. It also depends
on the
Num Seg Displayed
value you entered. For example, if you had entered
a value of 500 in
Num Segments
, and a value of 10 in
Num Seg Displayed
, the
maximum value you can enter as a starting segment is 491so that 10 segments
can be seen.