Quick Start 3
1.2 Power and Electrical Check
To ensure that the power supply will work according to it’s specifications,
please conduct an inspection before using the instrument.
Input Power Requirement
The power supply requires 50Hz / 60Hz power line frequency, at any of four AC
voltage levels: 100V / 120V / 220V / 230V, ±10%. Select the required power
voltage level with the “DIP Switch” on the rear panel.
Warning
Set the required input power voltage “DIP Switches” before
connecting the power cord.
Electrical Check
Please use the power cord provided and adjust the DIP Switches to the local
AC line voltage, then connect the instrument to AC power.
1. Set the DIP Switches to the local AC power voltage.
2. Connect the power cord to the T3PS3000 before connecting the power
cord to wall socket.
3. Use a DMM to check the power cord ground pin is connected to the
T3PS3000 ground terminal, then attach the power cord to the wall socket.
4. Turn on the power switch.
5. Press the button POWER to enter the T3PS3000 power on sequence. The
T3PS3000 will run it’s power on self tests then return to the power on
default settings.
Warning
To avoid electric shock, please make sure that the instrument
is
grounded correctly before powering on.