PS200
PORTABLE GAS MONITOR
CALIBRATION
64171, Rev. 11/December 2019
9-4
A. UPDATE: The monitor saves the calibration data in the memory and automatically updates the cali
-
bration due date based on the option stored in the monitor set-up menu.
B. SAVE ONLY: The monitor saves the calibration data in the memory but does not update the calibration
due date.
C. ABANDON: The monitor neither stores the calibration data in the memory nor updates the calibration
due date.
9. Press and hold the R button to accept this option.
9.3. Calibration Validity
Calibration validity remains the responsibility of the user. Individual codes of practice may dictate
calibration intervals.
Regular calibration establishes a pattern of reliability and enables the calibration check period to be
modified in line with operational experience. As a guide, the higher the risk, the more frequently cal
-
ibration should be checked.