PS200
PORTABLE GAS MONITOR
OPERATION
64171, Rev. 11/December 2019
3-2
• The oxygen sensor (if installed) operates correctly. The sensor responds to the user breathing
on the front of the monitor by displaying a value below 20.9%.
Additional:
• Perform regular leak checks on pumped instruments by placing your thumb over the sample
inlet nozzle and ensuring the monitor displays FLOW FAULT.
• Perform regular bump tests.
3.2. Quick Configuration Display
This allows configuration information to be viewed without fully switching the monitor ON (see
).
With the monitor OFF, press the R Button for less than one second. The monitor will display:
• sensors fitted;
• calibration due date;
• pumped or diffusion;
• LEL calibration gas.
Figure 3-2: Quick Configuration Display
3.3. Switching the Monitor On
Press and hold the R Button for one second to switch the monitor ON (refer to
). The flash screen appears (as shown in
) and the monitor begins
its warm-up routine. A countdown timer appears in the top right corner of the display.
WARNING: ALWAYS SWITCH THE MONITOR ON IN FRESH AIR. FAILURE TO HEED
THIS WARNING COULD RESULT IN SERIOUS INJURY OR DEATH.
NOTE: The LCD backlight is green during warm-up and automatically switches off when
warm-up is complete.
Содержание PS200 Series
Страница 10: ...64171 Rev 11 December 2019 x PS200 PORTABLE GAS MONITOR LIST OF TABLES This page is intentionally left blank ...
Страница 20: ...PS200 PORTABLE GAS MONITOR INTRODUCTION 64171 Rev 11 December 2019 2 4 This page is intentionally left blank ...
Страница 34: ...PS200 PORTABLE GAS MONITOR ALARMS 64171 Rev 11 December 2019 4 4 This page is intentionally left blank ...
Страница 54: ...PS200 PORTABLE GAS MONITOR ACCESSORIES 64171 Rev 11 December 2019 10 2 This page is intentionally left blank ...