PI-MAX
®
3 System Manual
Issue 2
7. Turn the CoolCUBE
II
on.
Verify there are no leaks or air bubbles in the hoses.
NOTE:
Small air bubbles about the size of bubbles in soda are
common in the CoolCUBE
II
particularly at start up and do
not prevent proper operation
•
If there are no problems, proceed to step 8.
•
If there are leaks or air bubbles:
a. Turn the CoolCUBE
II
off.
b. Correct any problem(s) by securing the hoses or adding more coolant to the
reservoir.
c. Turn the CoolCUBE
II
back on.
d. Recheck for leaks and/or air bubbles.
e. If there are no problems, proceed to step 8.
8. Turn on the PI-MAX3.
9. Launch the application software.
5.6
Connect PI-MAX3 to Dry Nitrogen Source
Equivalent Background Illumination (EBI) can impose limitations on ultra-low-light or
photon-counting applications. Reduction of EBI can be achieved by directly cooling the
photocathode via a dry nitrogen source.
Perform the following procedure to connect the PI-MAX3 to a dry nitrogen source:
1. Remove the 3 foot [0.9 m] clear PVC tubing from the Accessory Kit.
2. Slip one end of the tubing on the plastic hose barb on the front of the PI-MAX3.
3. Connect the other end to a source of dry nitrogen.
5.7
Configure Default System Parameters
This section provides information about configuring default system parameters in both
WinX/32 and LightField.
5.7.1
WinX/32 Applications (Version 2.5.25.X and Higher)
Perform the following procedure to configure default PI-MAX3 system parameters
when using WinX/32:
1. Verify the PI-MAX3 is connected to the host computer and that its power supply is
turned on.
2. Launch the WinX/32 application.
The Camera Detection Wizard will automatically run if this is the first time a
Teledyne Princeton Instruments application has been installed (i.e., WinView/32,
WinSpec/32, ore WinXTest/32,) and a supported camera is attached.
Otherwise, when installing a new camera type, click on the Launch Camera
Detection Wizard… on the Controller/CCD tab to start the wizard.
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...