Chapter 2
PI-MAX3 Camera System
25
Refer to
for information about each rear-panel connector and indicator.
Table 2-1:
PI-MAX3
Rear-Panel Connectors and Indicators (Sheet 1 of 2)
Label
Description
MONITOR
BNC port for TTL logic 1 pulse delayed ±3 ns with respect to
photocathode gating.
Cable delay, ~1.5 ns/ft, will be in addition to the delay at the connector.
I.I.T PWR ON/OFF
This switch biases the image intensifier photocathode ON or OFF.
• When the I.I.T. switch is set to ON, the photocathode can be gated
ON.
Exception: Selecting SAFE on the Experiment Setup Main screen
overrides control and will prevent the photocathode from being
biased on regardless of the I.I.T. switch setting.
• When the I.I.T. switch is set to OFF, then the photocathode cannot
be turned on from software.
CAUTION:
It is advisable to set the I.I.T. switch to OFF as a fail-safe
measure if the PI-MAX3 is left ON but unused for a period of
time.
ERROR
This RED LED lights to warn of excessive repetition rate.
Must be off for proper operation.
NOTE:
Excess Rep Rate LED is also activated by MCP bracket pulsing.
The MCP bracket repetition rate limit is 6.25 kHz.
TRIGGER IN
When external triggering is selected within the software, the internal
timing generator will be triggered by an externally derived trigger pulse
(range of ±5.5 V) applied to this input.
The threshold, slope, coupling mode (i.e., AC or DC,) and input
impedance (i.e., High or 50
,) are selectable in software.
The corresponding Green LED flashes each time PI-MAX3 is triggered. At
high repetition rates, the LED glows steadily.
NOTE:
In gated operation, the Green LED indicator flashes each time
the PI-MAX3 is triggered. The actual triggering can also be
readily determined by observing the signal at the PI-MAX3
Monitor output with a fast oscilloscope.
COOLANT PORTS:
• OUT
• IN
Two standard ¼” barbed brass fittings for circulating coolant are located
on the back panel.
Either port may be used as the inlet.
The coolant cannot be chilled.
Use the Teledyne Princeton Instruments CoolCUBE
II
coolant circulator.
For additional information, refer to:
•
Section 5.5, Connect PI-MAX3 to CoolCUBE
•
Section 7.5, Temperature Control
, on page 68.
NOTE:
Although circulating water will extend the PI-MAX3's cooling
capabilities, it is not necessary. Most of the PI-MAX3 cooling is
provided by the fan.
READY OUT
Level at this BNC is initially LOW.
It changes state on completion of cleaning cycles before the first exposure.
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...