1.5.1
Intensifier Modes and Safety
•
WinX/32 Applications
The Experiment Setup Main screen in WinX/32 applications allows you to select
one of two intensifier modes:
— Gate Mode;
In Gate Mode, the photocathode is biased on only for the time that each
gate pulse is applied. As a result, the tolerance to room light is higher in
gated operation, but the risk of damaging overload from intense light
sources such as lasers remains. In fact, intense light sources in gated
experiments can cause spot damage that would be undetected by the alarm
circuit.
— Safe Mode
In Safe Mode, the photocathode is continuously biased OFF and the
intensifier is as safe as it can be.
•
LightField
In LightField, you can enable or disable the intensifier on the Common
Acquisition Settings expander. When the intensifier is enabled, the camera can
be gated; when disabled, the photocathode is continuously biased OFF and the
intensifier is as safe as it can be.
NOTE:
In order for gating to occur, the I.I.T. switch on the back of
the PI-MAX3 must also be in the ON position.
1.5.2
Audible Alarm
NOTE:
It is normal for the alarm to sound briefly when the system is
turned on.
To reduce the risk of camera damage, the PI-MAX3 camera is equipped with an audible
alarm in the camera, activated when the intensity of light falling on the image intensifier
exceeds a preset threshold. While the alarm is sounding, the photocathode is disabled.
Immediately switch the I.I.T. switch on the back of the PI-MAX3 to the OFF position.
Cover the detector window and only switch the I.I.T. switch to ON after the illumination
level has been lowered. If the alarm sounds continuously even when the illumination
level is adequately low, shut the system down and contact the factory for guidance.
CAUTION!
!
Discontinue operation and contact the factory at once if
sporadic or continuous unwarranted alarms occur. They
may indicate intensifier damage or another situation that
requires immediate attention.
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...