154
PI-MAX
®
3 System Manual
Issue 2
3. Configure the PI-MAX3 to Gate Mode for the intensifier to operate properly.
This can be done by clicking the Gate Mode button on the Custom Toolbar, or by
selecting Gate Mode on the Acquisition
►
Experiment Setup…
►
Main tab. See
.
Figure 11-4: Typical Experiment Setup…
►
Main Tab
4. On the Acquisition
►
Experiment Setup…
►
Timing tab, verify Timing Mode is
configured for Single Trig. Mode. Continuous Cleans is also recommended for DIF.
See
.
Figure 11-5: Typical Experiment Setup…
►
Timing Tab
44
11
-0
12
9
_
0
10
3
441
1-
0
12
9
_
0
10
4
Содержание Princeton Instruments PI-MAX 3
Страница 1: ...www princetoninstruments com PI MAX 3 System Manual 4411 0129 Issue 2 August 20 2019 ...
Страница 32: ...32 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 38: ...38 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 124: ...124 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 150: ...150 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 174: ...174 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 178: ...178 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 190: ...190 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 214: ...214 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 236: ...236 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 244: ...244 PI MAX 3 System Manual Issue 2 This page is intentionally blank ...
Страница 249: ...This page is intentionally blank ...