94
NIRvana HS System Manual
1.0
E.3
External Coolant Circulator: Low Coolant (Air in the Hoses)
WARNING!
If more than 2” [50.8 mm] of the coolant line is filled with air,
the pump will stop working and may be damaged. If flow
stops while the pump is on, turn off the coolant circulator
and add coolant.
Perform the following procedure to troubleshoot a low coolant level:
1. Verify that the coolant circulator in question has been turned off, and its power
cord removed from the AC wall receptacle.
2. Remove the reservoir cap and verify that the coolant reservoir contains coolant.
If additional coolant is required, carefully add a 50:50 mixture of water and ethylene
glycol until it is at the recommended level.
3. Replace the reservoir cap.
4. Verify the power switch is in the off position.
5. Plug the circulator into an appropriate 100-240 V
AC
, 47-63 Hz power source.
6. Turn the circulator on.
7. Verify there are no leaks or air bubbles in the hoses.
NOTE:
Small air bubbles that are approximately the size of bubbles
in soda are common, particularly at start up, and do not
prevent proper operation.
•
If there are no problems, proceed to step 8.
•
If leaks or large air bubbles/pockets are observed, turn the circulator off and
attempt correct the problem(s) by securing the hoses or adding more coolant
to the reservoir. Turn the circulator back on.
— If there are no problems, proceed to step 8.
— If the problem persists, turn off the system, and contact Teledyne Princeton
Instruments Customer Support. Refer to
for complete contact information.
8. Turn the camera on.
9. Launch the application software.
Содержание Princeton Instruments NIRvana HS
Страница 1: ...www princetoninstruments com NIRvana HS System Manual 4411 0169 Issue 1 0 January 8 2020 ...
Страница 3: ...Issue 1 0 NIRvana HS System Manual 3 This page is intentionally blank ...
Страница 4: ...4 NIRvana HS System Manual Issue 1 0 ...
Страница 10: ...10 NIRvana HS System Manual Issue 1 0 ...
Страница 26: ...26 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Страница 34: ...34 NIRvana HS System Manual Issue 1 0 ...
Страница 52: ...52 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Страница 72: ...72 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Страница 90: ...90 NIRvana HS System Manual Issue 1 0 Figure D 1 Hg Spectrum NIRvana HS 1 7 SP 2500 4411 0169_0047 ...
Страница 91: ...Appendix D Calibration Charts 91 Figure D 2 Ne Ar Spectrum NIRvana HS 1 7 SP 2500 4411 0169_0048 ...
Страница 92: ...92 NIRvana HS System Manual Issue 1 0 This page is intentionally blank ...
Страница 101: ...This page is intentionally blank ...
Страница 106: ......