Automation
Section 4 The Column Selector Module
4-13
Figure 4-22 Connection of the waste tubing from the EZ
Prep inject valve to the left AutoInjector
Module port
16. Connect the supplied waste line (
P
/
N
60-5234-654) to the
right port of the AutoInjector Module.
Figure 4-23 Waste line connected to the right AutoInjector
Module port
17. Connect the check valve removed in Step 4 to the waste
line from the AutoInjector Module. Ensure that the arrow
on the check valve points away from the AutoInjector Mod-
ule as shown in Figure 4-24.