Automation
Section 3 The AutoSampler Module
3-20
3.3 AutoSampler Module
Operation
The AutoSampler Module increases the fraction collection
capacity of the CombiFlash EZ Prep by two racks. The RFID
capabilities of these additional racks are supported in the same
way as the racks for the EZ Prep. When a rack is full, the system
will automatically advance to the next available rack. If a rack is
replaced with a fresh rack, this will be sensed by the system
software and the new rack will be made available for use. The
sample racks also have RFID tags which allow for the recognition
of vial locations.
Note
The sample rack cannot be swapped during a separation
sequence.
3.3.1 Setup
Before using the AutoSampler Module for the first time, the EZ
Prep should be configured for the proper injection loop volume.
1. Begin by going to T
OOLS
| C
ONFIG
| P
REP
HPLC.
2. Select the loop volume field as shown in Figure 3-38.
Note
The screen shot in Figure 3-38 depicts what is shown when a
Column Selector Module is installed. If the Column Selector
Module is not installed, this screen may appear differently.
Figure 3-38 Configuring the injection loop volume
Содержание AutoInjector Module
Страница 2: ......
Страница 6: ...Automation Modules Installation and Operation Guide Safety Warnings 6 ...
Страница 8: ...Automation Modules Table of Contents 8 ...
Страница 10: ...Automation Modules Section 1 Introduction 1 2 ...
Страница 46: ...Automation Section 3 The AutoSampler Module 3 24 ...
Страница 64: ...Automation Section 4 The Column Selector Module 4 18 ...
Страница 68: ...Automation Appendix A Specifications A 4 ...
Страница 69: ...Compliance Statements C E D e c l a r a t i o n o f C o n f o r m i t y P e n d i n g ...
Страница 70: ......
Страница 72: ......
Страница 74: ......