Automation
Section 2 The AutoInjector Module
2-6
Figure 2-9 Waste line connected to the right AutoInjector
Module port
17. Connect the check valve removed in Step 4 to the waste
line from the AutoInjector Module. Ensure that the arrow
on the check valve points away from the AutoInjector Mod-
ule as shown in Figure 2-10.
Figure 2-10 Check valve connected to the waste line of the
AutoInjector Module
18. Locate the supplied sample probe tubing and fitting pack-
age (
P
/
N
60-5234-657).
19. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
20. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.