Figure 3-16 Gap between the Z-axis slider and the cap at
the top of the Z-Drive
21. Hold the cable flat against the rotor and secure the cable
by tightening the thumbscrew on the rotor. The thumb-
screw should be as tight as possible using just your fingers.
Figure 3-17 Probe cable secured onto the rotor
22. Rotate the Z-Drive rotor back and forth to ensure that the
Z-Drive moves up and down freely.
Note
If the rotor does not move freely, check that the other cables
are not interfering with the movement of the green probe cable.