Automation
Section 2 The AutoInjector Module
2-3
Figure 2-4 AutoInjector Module attached to the EZ Prep
column mount
6. Connect the USB cable from the AutoInjector Module to
the USB port on the rear of the EZ Prep. If you have a
CombiFlash Rf+ PurIon
TM
Mass Detector attached to the
EZ Prep, you will need to obtain a USB hub to connect both
modules.
7. Connect the included power supply from the AutoInjector
Module to a suitable power connection. Use a power supply
in a location that minimizes the possibility of a chemical
spill damaging the power supply.
8. Locate the supplied waste tubing and fitting package (
P
/
N
60-5234-653).
9. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
10. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.