Automation
Section 4 The Column Selector Module
4-14
Figure 4-24 Check valve connected to the waste line of the
AutoInjector Module
18. Locate the supplied sample probe tubing and fitting pack-
age (
P
/
N
60-5234-657).
19. Unscrew the attached plastic ferrule in this fitting package
until the plastic ferrule is almost completely detached from
the steel fitting.
20. Attach the steel fitting in this package to the EZ Prep
inject valve waste port. This steel fitting should be hand
tight.
Figure 4-25 Sample tube fitting attached to the EZ Prep
inject valve
21. Turn the plastic ferrule until it is attached hand tight to
the valve.
22. Turn the steel fitting
1
/
2
turn tighter.
23. Remove the plastic ferrule while leaving the fitting
attached.