Operation
Model 3300TB
Teledyne Analytical Instruments
26
4.7 Displaying Percent & PPM on the LED
Display
The analyzer displays the concentration in percent whenever the
reading is over 9999 ppm. When the reading changes to percent, the
LED display will alternate between flashing ―PC‖ and the oxygen
concentration. On the other hand, if the instrument is displaying ppm,
only the concentration reading will be shown.
4.8 Supplementary Information
If, during the Span Procedure, you pressed the SPAN button by
mistake, you must wait five minutes for the analyzer to resume analisis
or you can press the UP button and then the DOWN button. (Pressing
the UP and DOWN buttons causes the analyzer to time-out in five
seconds instead of five minutes).
If during the span procedure, you press the RANGE or ALARM
buttons:
- either the range or alarm routine will be activated.
- any changes to span will be rejected.
- the 60 second alarm delay will not occur, i.e., the alarms will be
responsive immediately.
CAUTION: TAI considers the action of pressing the ALARM
or
RANGE buttons to be at your (the user’s)
discretion. The user should be aware that the
alarms will become active within 5 (five) seconds
if the Range or Alarm bottons are pressed during
Span. This may result in false alarms if the Span
gas has not been fully purged from the Cell and
Sample lines.