Trace Oxygen Analyzer
Operational Theory
Teledyne Analytical Instruments
11
The output from the temperature compensation amplifier is sent to
an analog to digital converter (ADC), and the resulting digital
concentration signal is sent to the microcontroller.
The digital concentration signal along with input from the front
panel buttons (KEYBOARD) is processed by the microcontroller, and
appropriate output signals are directed to the display and alarm relays.
The same digital information is also sent to a 12-bit digital to analog
converter (DAC) that produces the 0-10 V dc analog concentration
signal and the 0-10 V dc analog range ID output. A voltage current
converter (E–I CONV) produces the 4-20 mA dc analog concentration
signal.
Содержание 3300TB