Trace Oxygen Analyzer
Installation
Teledyne Analytical Instruments
19
Select Sample with the three-way control valve and open the flow
control valve on the analyzer to full flow (located on the analyzer flow
meter). Open the pump by-pass valve, and turn on the pump. Readjust
the pump by-pass valve to limit the maximum flow rate to full scale on
the flowmeter with the flowmeter valve fully open. Adjust the analyzer
flow control valve (flowmeter valve) to a flow-rate of approximately 2
SCFH). Select Span and verify that the span and Sample gas are at
different pressures. The calibration accuracy will be adversely affected
if the Span and Sample gases are at differing pressures.
If the span gas is provided from pressurized source, a control valve
must be added between the analyzer and the span gas source. This valve
is used to adjust the span flow rate to match the sample flow rate. Do
not adjust the span flow rate with the flow control valve located on the
analyzer for units with vacuum service options.
Figure 3-4: Piping Diagram for Vacuum Service Option
3.5 Installation Checklist
Before connecting the instrument to the power source and turning it
on, make sure you have:
Correctly installed the Sample and Exhaust gas lines
Checked for leaks