Trace Oxygen Analyzer
Installation
Teledyne Analytical Instruments
13
Installation
Installation of the Model 3300TB Analyzer includes:
1.
Unpacking the system.
2.
Mounting the Control Unit to a wall
3.
Installing the Micro-Fuel Cell
4.
Making the electrical connections.
5.
Making the gas connections.
6.
Testing the installation.
CAUTION: Read this chapter in its entirety before installing
the units.
The Sample must be free of entrained solids or
water. However, a high humidity sample is ideal,
since it will prevent water loss from the cell
electrolyte.
The Micro-Fuel Cell sensor electrolyte is caustic.
Do not attempt to open it. Leaking or exhausted
cells should be disposed of in accordance with
local regulations. Refer to the Material Safety
Data Sheet in the Appendix.
Any damage or scarring of the delicate
permeable membrane on the sensing end of the
cell will require cell replacement. Prevent contact
with membrane by any solid object.
3.1 Unpacking the Analyzer
As soon as you receive the instrument, carefully unpack and inspect
the Unit, and any included accessories for damage. Immediately report
any damage to the shipping agent. The analyzer is shipped with all the
materials you need to install and prepare the system for operation.
CAUTION: Do not disturb the integrity of the cell package
until the cell is to actually be used. If the cell
package is punctured and air is permitted to
enter, cell-life will be compromised.