
Trace Oxygen Analyzer
Installation
Teledyne Analytical Instruments
15
During the installation and/or replacement of the MFC, the
membrane surface MUST ALWAYS FACE
DOWNWARD, and the
contact side of the membrane, must be placed FIRST into Analysis Unit.
The reason for proper installation/replacement is, if any bubble that
develops as the electrolyte dries out it will be directed by gravity away
from the membrane.
To install or replace the Micro-Fuel Cell, follow the procedures in
Chapter 5,
Maintenance
.
3.3 Electrical Connections
Figure 3-1 shows the Model 3300TB rear panel of the control unit.
For detailed pinouts, see the wiring/interconnection drawings in the
Drawings section at the rear of this manual. To gain access to the
terminal blocks, the analyzer door screws must be loosened, and the
door opened.
Figure 3-1 Electrical Connectors for AC Control Unit
Primary Input Power:
The power strip is installed inside the
analyzer. Connect ground to terminal 1, neutral to terminal 3, and hot to
terminal 2. Make sure female plug end is inserted in the control unit
power receptacle.
Содержание 3300TB
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 25: ...Operational Theory Model 3300TB Teledyne Analytical Instruments 12 Blank Page ...
Страница 45: ...Maintenance Model 3300TB Teledyne Analytical Instruments 32 ...
Страница 51: ......