CP030 Current Probe
20
922173-00 Rev A
Appendix A
Performance Verification Test Record
Photocopy this page and record the results of measurements made during the performance verification
of the CP030 Current Probe on the copy. File the completed record as required by applicable internal
quality procedures.
The sections in the test record correspond to the parameters tested in the performance verification
procedure. The numbers preceding the individual data records correspond to the steps in the procedure
that require the recording of data. Results recorded in the column labeled "Test Result" are the actual
specification limit check. The test limits are included in all of these steps. Other measurements and the
results of intermediate calculations that support the limit check are to be recorded in the column
labeled "Intermediate Results".
Permission is granted to reproduce these pages for the purpose of recording test results.
Model:
_________
________
______
Serial Number:
_________
________
______
Asset or Tracking Number:
_________
________
______
Date:
_________
________
______
Technician:
_________
________
______
Equipment Used
MODEL
SERIAL NUMBER
CALIBRATION DUE DATE
Digital Multimeter #1
Digital Multimeter #2
Function Generator
1
N/A
1
The function generator used in this Performance Verification Procedure is used for making relative
measurements. The output of the generator is measured with a DMM or oscilloscope in this procedure.
Thus, the generator is not required to be calibrated.
CP030 Test Record
Step
Description
Test Result
Gain Accuracy
10 Probe Output (Spec limit: 0.495 - 0.505 V)
________________ V