CP030 Current Probe
Instruction Manual
January 2013
www.calcert.com
[email protected]
1.888.610.7664
0
5
10
15
20
25
30
Страница 1: ...Instruction Manual CP030 Current Probe www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30 ...
Страница 2: ...CP030 Current Probe Instruction Manual January 2013 www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30 ...
Страница 3: ...cising its warranty Teledyne Lecroy at its option will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center However this will be done only if the product is determined by Teledyne Lecroy s examination to be defective due to workmanship or materials and the defect is not caused by misuse neglect accident abnorma...
Страница 4: ...t 4 Connecting the Probe to the Test Circuit 4 Operation with a Teledyne LeCroy Oscilloscope 5 Bandwidth Limit 5 Auto Zero 5 Degauss Probe 5 Care and Maintenance 6 Cleaning 6 Calibration Interval 6 Service Strategy 6 Troubleshooting 6 Returning a Defective Probe 7 Performance Verification 8 Test Equipment Required 8 Preliminary Procedure 9 Functional Check 9 Check LF Accuracy 10 www calcert com sa...
Страница 5: ...tics 15 Graphs 16 Certifications 17 EMC Compliance 17 EC Declaration of Conformity EMC 17 Australia New Zealand Declaration of Conformity EMC 17 Safety Compliance 18 EC Declaration of Conformity Low Voltage 18 Environmental Compliance 18 End Of Life Handling 18 Restriction of Hazardous Substances RoHS 18 Contact Teledyne LeCroy 19 Appendix A 20 Performance Verification Test Record 20 Equipment Use...
Страница 6: ...t personal injury or damage Do not proceed until conditions are fully understood and met ELECTROSTATIC DISCHARGE ESD HAZARD The probe is susceptible to damage if anti static measures are not taken DOUBLE INSULATION PROTECTIVE EARTH TERMINAL Precautions Connect and disconnect properly Connect probe to the measurement instrument before connecting the test leads to a circuit signal being tested Use o...
Страница 7: ...g measured carry less than 300 V but more than the safe voltage level SELV E make sure that the conductors to be measured are insulated with material conforming to o Overvoltage Category I basic insulation requirement for working voltage of 300 V o Pollution degree 2 Operating Environment The accessory is intended for indoor use and should be operated in a clean dry environment Before using this p...
Страница 8: ... with firmware 4 3 0 0 or higher With the ProBus interface the CP030 becomes an integral part of the oscilloscope The bandwidth limit auto zero and degauss functions are all controlled from the oscilloscope s graphical user interface The oscilloscope provides power to the probe so no external power supply is needed Key Features Highly accurate current measurements Easy current measurements Wide ba...
Страница 9: ... Attach the probe output connector to the oscilloscope input connector The oscilloscope will recognize the probe set the oscilloscope input termination to 1 MΩ and activate the probe control functions in the user interface To use the CP030 with a WaveMaster scope the AP 1M impedance adapter is required Connecting the Probe to the Test Circuit The CP030 has been designed with a movable split core e...
Страница 10: ...ero The CP031 incorporates an Auto Zero function to remove the DC offset from the current probe Auto Zero must be invoked by the user After several minutes of warm up or when the probe is exposed to a large shift in ambient temperature some DC offset drift may occur To initiate an Auto Zero cycle touch the CP031 tab of the Channel Vertical Setup dialog then touch AUTO ZERO Degauss Probe If the pro...
Страница 11: ...e probe under warranty will be replaced with a factory refurbished probe A probe that is not under warranty can be exchanged for a factory refurbished probe A modest fee is charged for this service A defective probe must be returned in order to receive credit for the probe core Troubleshooting If the probe is not operating properly the problem may be the way in which it is used Before assuming the...
Страница 12: ...nded by the original packing material or equivalent and box 4 Label the case with a tag containing The RAN Name and address of the owner Probe model and serial number Description of failure 5 Package the probe case in a cardboard shipping box with adequate padding to avoid damage in transit 6 Mark the outside of the box with the shipping address given to you by the Teledyne Lecroy representative b...
Страница 13: ...ns required to meet the accuracy needed for the procedure Make sure that the ambient conditions meet the requirements of all the test instruments used in his procedure NOTE The correct operation of the CP030 controls requires software version 4 3 0 0 or higher The software version in the test oscilloscope can be verified by selecting Utilities Utilities Setup Status tab Contact your local Teledyne...
Страница 14: ...warmup time for the CP030 and test equipment before performing the Verification Procedure 3 Turn on the other test equipment and allow these to warm up for the time recommended by the manufacturer 4 While the instruments are reaching operating temperature make a photocopy of the Performance Verification Test Record located in Appendix A and fill in the necessary data Functional Check The functiona...
Страница 15: ...the Current Sense terminals of the 100 Turn Calibration Loop to the voltage inputs of DMM 2 5 Set both DMMs to measure AC Volt 6 With the CP030 removed from any signal and the slider returned to the LOCKED position degauss the probe by pressing the DEGAUSS button located in the CP030 dialog then OK 7 Open the CP030 slider and position the probe input around the 100 Turn loop Close and LOCK the sli...
Страница 16: ...act your local Teledyne Lecroy representative if the software in your oscilloscope requires updating Test Equipment Required The following table lists the test equipment and accessories or their equivalents which are required for adjustment of the CP030 Current Probe Because the input and output connector types may vary on different brands and models of test instruments additional adapters or cabl...
Страница 17: ...rd out of the box 2 Connect the CP030 compensation board to the channel 1 input of the oscilloscope through the ProBUS extension cable and completely close the probe slider 3 Turn the oscilloscope on and allow at least 30 minutes warmup time for the CP030 and test equipment before performing the Verification Procedure Adjust LF Accuracy Figure 2 LF accuracy adjustment set up 1 Set the Function Gen...
Страница 18: ...lloscope located in the CP030 dialog then pressing OK 7 Open the CP030 slider and position the probe input around the 100 Turn loop Close and LOCK the slider 8 Adjust the Function Generator voltage until the voltage measured at the Current Sense terminals DMM 2 is 50 mV 0 05 mV This corresponds to 10 A at the probe head 9 Adjust VR202 on the PCB until the voltage measured on DMM 1 is as close to 2...
Страница 19: ... specifications Guaranteed at 23 C 5 C 73 F 9 F after power has been applied for 30 minutes Amplitude Accuracy 1 0 of reading 10 mA to 30 Arms 2 0 of reading to 50 Apeak DC 45 to 65 Hz Typical Characteristics Typical characteristics describe parameters which do not have guaranteed performance however are representative of the average performance from a sample of several probes Tests for typical ch...
Страница 20: ... to 2000 m 6562 feet Effect of External Magnetic Equivalent to a maximum of Field 20 mA In a DC or 60 Hz 400 A m magnetic field Physical Characteristics Dimensions Probe Length 175 mm 6 9 inch Width 18 mm 0 7 inch Height 40 mm 1 57 inch Dimensions Compensation Box Length 65 mm 2 6 inch Width 39 mm 1 5 inch Height 24 mm 0 9 inch Weight 240 g 8 5 oz Maximum diameter of conductors 5 mm 0 2 inch to be...
Страница 21: ... Current Probe 16 922173 00 Rev A Graphs Figure 3 Maximum input current vs frequency Frequency Hz Figure 4 Insertion impedance vs frequency www calcert com sales calcert com 1 888 610 7664 0 5 10 15 20 25 30 ...
Страница 22: ...EN 61000 4 3 2006 RF Radiated Electromagnetic Field 3 V m 80 1000 MHz 3 V m 1400 MHz 2 GHz 1 V m 2 GHz 2 7 GHz 3 1 Emissions which exceed the levels required by this standard may occur when the probe is connected to a test object 2 This product is intended for use in nonresidential areas only Use in residential areas may cause electromagnetic interference 3 Meets Performance Criteria B limits of t...
Страница 23: ...asuring circuits EN 61010 031 A1 2008 Safety requirements for electrical equipment for measurement control and laboratory use Part 031 Safety requirements for hand held probe assemblies for electrical measurement and test Environmental Compliance End Of Life Handling The instrument is marked with this symbol to indicate that it complies with the applicable European Union requirements to Directives...
Страница 24: ... calculations that support the limit check are to be recorded in the column labeled Intermediate Results Permission is granted to reproduce these pages for the purpose of recording test results Model _______________________ Serial Number _______________________ Asset or Tracking Number _______________________ Date _______________________ Technician _______________________ Equipment Used MODEL SERI...