Instruction Manual
922173-00 Rev A
5
Operation with a Teledyne LeCroy Oscilloscope
The probe can be controlled through the Vertical Setup dialog for the channel to which the probe is
connected.
Turning the
VOLTS/DIV
knob will control the oscilloscope’s scale factor to give full dynamic range from
20 mA/div to 50 A/div.
The CP030 probe dialog (a tab behind the channel setup dialog) allows for the selection of the probe’s
coupling (
DC, Grounded
or
AC
),
AUTO ZERO, DEGAUSS PROBE
and
Probe BWL
functions, limiting the
system bandwidth to 20 MHz or maximum bandwidth (BWL Off).
Bandwidth Limit
The CP030 is capable of switching the bandwidth from
Full
(maximum bandwidth) to
20 MHz
by
selecting the bandwidth adjustment in the channel menu.
Auto Zero
The CP031 incorporates an Auto Zero function to remove the DC offset from the current probe. Auto
Zero must be invoked by the user. After several minutes of warm-up, or when the probe is exposed to a
large shift in ambient temperature, some DC offset drift may occur. To initiate an Auto Zero cycle, touch
the CP031 tab of the (Channel) Vertical Setup dialog, then touch
AUTO ZERO
.
Degauss Probe
If the probe has been magnetized by external magnetic field or by excessive input, the core can be
demagnetized by opening the CP030 dialog and touching
DEGAUSS PROBE
. The demagnetizing process
takes about 5 seconds and should always be performed before taking a measurement.
Without clamping the probe around a conductor, slide the opening lever to close and lock the probe and
press the
DEGAUSS
PROBE
button.
An Auto Zero is automatically performed as part of the degauss cycle.