44
Teledyne API Model 480L O
3
Monitor User Manual
09419A DCN8342
Discrete Input Registers (Read Only)
Register
Address
Description
Tag Name
Data Type
# Bits
6
Stream 1 HI-HI Alarm
Status, 1= ON
STREAM_1_HI_HI_ALARM Boolean
1
7
Stream 2 HI-HI Alarm
Status, 1= ON
STREAM_2_HI_HI_ALARM Boolean
1
8
Stream 3 HI-HI Alarm
Status, 1= ON
STREAM_3_HI_HI_ALARM Boolean
1
9
Stream 4 HI-HI Alarm
Status, 1= ON
STREAM_4_HI_HI_ALARM Boolean
1
10
Stream 5 HI-HI Alarm
Status, 1= ON
STREAM_5_HI_HI_ALARM Boolean
1
11
Stream 6 HI-HI Alarm
Status, 1= ON
STREAM_6_HI_HI_ALARM Boolean
1
12
Stream 1 Active Status,
1= Sampling
STREAM_1_ACTIVE
Boolean
1
13
Stream 2 Active Status,
1= Sampling
STREAM_2_ACTIVE
Boolean
1
14
Stream 3 Active Status,
1= Sampling
STREAM_3_ACTIVE
Boolean
1
15
Stream 4 Active Status,
1= Sampling
STREAM_4_ACTIVE
Boolean
1
16
Stream 5 Active Status,
1= Sampling
STREAM_5_ACTIVE
Boolean
1
17
Stream 6 Active Status,
1= Sampling
STREAM_6_ACTIVE
Boolean
1
6.3.4 INSTRUMENT SETUP AND CONFIGURATION REGISTERS (READ /
WRITE)
Coil Registers (Read/Write)
Register
Address
Description
Tag Name
Data
Type
# Bits
0
Stream 1 Enable,
1=ON
STREAM_1_ENABLE
Boolean
1
1
Stream 2 Enable,
1=ON
STREAM_2_ENABLE
Boolean
1
2
Stream 3 Enable,
1=ON
STREAM_3_ENABLE
Boolean
1
3
Stream 4 Enable,
1=ON
STREAM_4_ENABLE
Boolean
1
4
Stream 5 Enable,
1=ON
STREAM_5_ENABLE
Boolean
1
5
Stream 6 Enable,
STREAM_6_ENABLE
Boolean
1