Performance Verification
2- 48
WFM700 Series Waveform Monitors Service Manual
13.
While observing waveform amplitude, adjust the RF generator frequency
down to 500 MHz (in 10 MHz steps). If the waveform amplitude falls below
the cursors set in the previous step (step 12), re-adjust the cursors to the new
minimum level.
14.
Calculate the high frequency response using the formula:
High Frequency Response
20
log
10
Æ
Measured Minimum
800 mV
É
15.
CHECK
that the calculated high frequency response is between 0 and –3 dB.
16.
Set the RF generator frequency to 50 kHz.
17.
Measure the waveform amplitude using the cursors and note the measure-
ment.
18.
Calculate the low frequency response using the formula:
Low Frequency Response
20
log
10
Æ
Measured Minimum
800 mV
É
19.
CHECK
that the calculated low frequency response is between 0 and –3 dB.
20.
Move the input cable assembly from
Input A
to
Input B
.
21.
Press the
CURSOR
button to exit cursor mode.
22.
Touch the
Input#
soft key, touch
Input B
, then touch
Done
.
23.
Repeat steps 7 through 19 for
Input B
.
24.
Press the
CURSOR
button to exit cursor mode.
25.
Touch
Exit
to exit Eye Bandwidth mode.
NOTE
. Most front panel buttons are disabled while in the Eye Sig BW Test mode.
You must use the Exit soft key to return instrument functions.
26.
Disconnect the test setup.
Содержание WFM700 Series
Страница 4: ......
Страница 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Страница 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Страница 19: ...Theory of Operation...
Страница 20: ......
Страница 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Страница 31: ...Performance Verification...
Страница 32: ......
Страница 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Страница 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Страница 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Страница 107: ...Adjustment Procedures...
Страница 108: ......
Страница 131: ...Adjustments WFM700 Series Waveform Monitors Service Manual 3 23 Figure 3 14 Jitter HPF adjustment screen...
Страница 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Страница 133: ...Maintenance...
Страница 134: ......
Страница 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Страница 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Страница 177: ...Replaceable Parts...
Страница 178: ......
Страница 193: ...Replaceable Parts WFM700 Series Waveform Monitors Service Manual 5 15 1 2 3 4 5 Figure 5 4 WFM700 Audio board option DG...
Страница 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...