Performance Verification
2- 34
WFM700 Series Waveform Monitors Service Manual
4.
Press the
INPUT
button and select
Input A
.
5.
Press the
EYE
button and then touch the
Display
soft key.
6.
Touch the
Jitter
soft key and then touch the
Done
soft key.
7.
Touch the
Hi Pass Filter:
soft key and set to
100 kHz
.
8.
CHECK
that the Measured Jitter is approximately 30 ps (typical).
9.
Touch the
Hi Pass Filter:
soft key and set to
10 kHz
.
10.
CHECK
that the Measured Jitter is approximately 60 ps (typical).
11.
Change the RF generator settings as follows:
Frequency
135.0 MHz
12.
Touch the
Hi Pass Filter:
soft key and set to
10 Hz
.
13.
CHECK
that the Measured Jitter is approximately 200 ps (typical).
This test checks the Eye Time Base Jitter.
Typical Operation Check.
This test checks for typical operation. Typical values are
not guaranteed characteristics and are not listed in the test record.
NOTE
. If this test is performed out of sequence or the sequence of tests has been
interrupted (such as power off or input from another module was selected),
perform the steps in this note to prepare the instrument to continue testing.
a.
Connect the SD video generator to
INPUT B
using a 75
8
cable.
b.
Set the SD video generator for a 525-format 100% color bar signal.
c.
Press the
INPUT
button and select
Input B
.
d.
Press the
EYE
button.
1.
Install the 50
8
N-to-BNC adapter on the RF generator output.
2.
Connect the RF generator output to
INPUT A
using a 75
8
cable. Imped-
ance matching is not important for this test.
3.
Set the RF generator as follows:
Frequency
741.7582 MHz
Output Level
0 dB
Frequency Modulation
OFF
Phase Modulation
OFF
RF Output
ON
Eye Time Base Jitter
(WFM7M only)
Содержание WFM700 Series
Страница 4: ......
Страница 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Страница 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Страница 19: ...Theory of Operation...
Страница 20: ......
Страница 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Страница 31: ...Performance Verification...
Страница 32: ......
Страница 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Страница 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Страница 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Страница 107: ...Adjustment Procedures...
Страница 108: ......
Страница 131: ...Adjustments WFM700 Series Waveform Monitors Service Manual 3 23 Figure 3 14 Jitter HPF adjustment screen...
Страница 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Страница 133: ...Maintenance...
Страница 134: ......
Страница 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Страница 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Страница 177: ...Replaceable Parts...
Страница 178: ......
Страница 193: ...Replaceable Parts WFM700 Series Waveform Monitors Service Manual 5 15 1 2 3 4 5 Figure 5 4 WFM700 Audio board option DG...
Страница 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...