Adjustments
WFM700 Series Waveform Monitors Service Manual
3- 21
This procedure sets the calibration constant for the eye diagram clock recovery
bandwidth. This calibration routine is only available for the WFM700M
(WFM700 with a WFM7M module).
Following is a list of the signal requirements to perform the adjustment.
Adjacent to the requirement is the recommended equipment and test signals.
Adjustment requirements
Recommended test equipment and test signal
Calibrated sine wave generator,
741.7582 MHz, 800 mV into
75
8
, with 1 kHz jitter
modulation
With the recommended RF generator, you can generate the
required signal with the following settings to obtain 1 kHz
jitter modulation.
Frequency
741.7582 MHz
Output Level
0 dB
LFGEN2 Frequency
1 kHz
LFGEN2 Waveshape
SINE
Phase Modulation Source
LFGEN2
Phase Deviation
2.22 RADIANS
RF Output
ON
Phase Modulation
ON
75
8
coaxial cable
36 inches long
NOTE
. If this test is performed out of sequence or the sequence of tests has been
interrupted (such as power off or input from another module was selected),
perform the steps in this note to prepare the instrument to continue testing.
a.
Connect the SD video generator to
INPUT B
using a 75
8
cable.
b.
Set the SD video generator for a 525-format 100% color bar signal.
c.
Press the
INPUT
button and select
Input B
.
d.
Press the
EYE
button.
e.
Press the
INPUT
button and select
Input A
.
1.
Install the 50
8
N-to-BNC adapter on the RF generator output.
2.
Connect the RF generator output to
INPUT A
using a 75
8
cable. Imped-
ance matching is not important for this connection.
3.
Touch the
Eye Loop BW Adjust (1 kHz)
soft key and then the
Start Eye
Loop BW Adjustment (1 kHz)
soft key.
4.
Adjust the RF generator output level for a displayed waveform amplitude of
approximately 800 mV.
5.
Adjust the displayed modulation using the General Purpose Knob for exactly
1 UI as measured on the display. See Figure 3--13 on page 3--18.
6.
Touch the
Save and Exit
soft key to save the new calibration constant.
Eye Loop BW Adjust
(1 kHz)
Содержание WFM700 Series
Страница 4: ......
Страница 10: ...Table of Contents vi WFM700 Series Waveform Monitors Service Manual...
Страница 14: ...Service Safety Summary x WFM700 Series Waveform Monitors Service Manual...
Страница 19: ...Theory of Operation...
Страница 20: ......
Страница 30: ...Theory of Operation 1 10 WFM700 Series Waveform Monitors Service Manual...
Страница 31: ...Performance Verification...
Страница 32: ......
Страница 88: ...Performance Verification 2 56 WFM700 Series Waveform Monitors Service Manual...
Страница 94: ...WFM700 WFM7M Input Module Test Record 2 62 WFM700 Series Waveform Monitors Service Manual...
Страница 102: ...WFM700 WFM7HD Input Module Test Record 2 70 WFM700 Series Waveform Monitors Service Manual...
Страница 107: ...Adjustment Procedures...
Страница 108: ......
Страница 131: ...Adjustments WFM700 Series Waveform Monitors Service Manual 3 23 Figure 3 14 Jitter HPF adjustment screen...
Страница 132: ...Adjustments 3 24 WFM700 Series Waveform Monitors Service Manual...
Страница 133: ...Maintenance...
Страница 134: ......
Страница 160: ...Removal and Installation Procedures 4 26 WFM700 Series Waveform Monitors Service Manual...
Страница 176: ...Repackaging Instructions 4 42 WFM700 Series Waveform Monitors Service Manual...
Страница 177: ...Replaceable Parts...
Страница 178: ......
Страница 193: ...Replaceable Parts WFM700 Series Waveform Monitors Service Manual 5 15 1 2 3 4 5 Figure 5 4 WFM700 Audio board option DG...
Страница 194: ...Replaceable Parts 5 16 WFM700 Series Waveform Monitors Service Manual...