Performance Verification and Adjustment Procedures
TSG 95 Service Manual
4–7
NOTE
. After initial delivery or long storage, allow a two-hour warm-up time to
re-age the crystal. Thereafter, a 20 minute warm-up time is sufficient.
1. Check Subcarrier Frequency
a. Connect the TSG 95 Video output to the genlock input of the NTSC test
signal generator with a 75
Ω
coaxial cable as shown in Figure 4–2.
Terminate the genlock input in 75
Ω
, if required.
b. Connect the NTSC test signal generator subcarrier output to the frequen-
cy counter A input, using a 75
Ω
coaxial cable.
c.
Connect a reference, such as WWV, to the B input of the frequency
counter.
d. Set the frequency counter for a ratio of A/B, and set the AVG to 106.
e.
If the TSG 95 is not set to generate NTSC signals perform the following
steps, otherwise continue on at step f.
H
Press and hold the Lock Out key down, then press the ON key.
H
Select the NTSC video standard by pressing either the
A
or
"
key
until the standard is displayed.
H
Press the
Y
key to move to EXIT MENU, then press Enter to leave
the Utility menu.
H
To change the VM700A video standard, press the Measure and
<Video Standard> keys.
f.
CHECK — that the measured oscillator frequency is
3.579545 MHz
±
10 Hz.
g. Repeat steps a through e for PAL by substituting “PAL” where ever
“NTSC” is called out.
h. CHECK — that the measured oscillator frequency is
4.43361875 MHz
±
10 Hz.
Oscillator Frequency
Содержание TSG 95
Страница 4: ......
Страница 6: ...TSG 95 Service Manual...
Страница 14: ...Service Safety Summary viii TSG 95 Service Manual...
Страница 29: ...Specifications TSG 95 Service Manual 1 15 Figure 1 3 PAL 75 Red and Red Field Figure 1 4 PAL 100 Red...
Страница 30: ...Specifications 1 16 TSG 95 Service Manual Figure 1 5 PAL Green Field Figure 1 6 PAL Blue Field...
Страница 31: ...Specifications TSG 95 Service Manual 1 17 Figure 1 7 PAL 50 Flat Field Figure 1 8 PAL 100 Flat Field...
Страница 34: ...Specifications 1 20 TSG 95 Service Manual Figure 1 13 PAL Modulated 5 Step Figure 1 14 PAL Multiburst...
Страница 35: ...Specifications TSG 95 Service Manual 1 21 Figure 1 15 PAL Reduced Sweep Figure 1 16 PAL Pluge...
Страница 40: ...Specifications 1 26 TSG 95 Service Manual Figure 1 25 PAL Matrix Signal UK ITS 2...
Страница 41: ...Specifications TSG 95 Service Manual 1 27 Figure 1 26 PAL Safe Area...
Страница 42: ...Specifications 1 28 TSG 95 Service Manual NTSC Waveform Diagrams Figure 1 27 SMPTE NTSC Color Bar Components...
Страница 43: ...Specifications TSG 95 Service Manual 1 29 Figure 1 28 SMPTE NTSC Color Bars Zero Setup...
Страница 45: ...Specifications TSG 95 Service Manual 1 31 Figure 1 31 NTSC Red Field Figure 1 32 NTSC Red Field Zero Setup...
Страница 46: ...Specifications 1 32 TSG 95 Service Manual Figure 1 33 NTSC Sin x x Figure 1 34 NTSC 5 Step Staircase Gray Scale...
Страница 50: ...Specifications 1 36 TSG 95 Service Manual Figure 1 41 NTSC Convergence Components...
Страница 51: ...Specifications TSG 95 Service Manual 1 37 Figure 1 42 NTC7 NTSC Composite Figure 1 43 NTC7 NTSC Combination...
Страница 52: ...Specifications 1 38 TSG 95 Service Manual Figure 1 44 FCC NTSC Composite Figure 1 45 NTSC Cable Multiburst...
Страница 53: ...Specifications TSG 95 Service Manual 1 39 Figure 1 46 NTSC Cable Sweep Figure 1 47 NTSC Matrix...
Страница 54: ...Specifications 1 40 TSG 95 Service Manual Figure 1 48 NTSC Safe Area...
Страница 104: ...Performance Verification and Adjustment Procedures 4 30 TSG 95 Service Manual...
Страница 118: ...Maintenance 5 14 TSG 95 Service Manual...
Страница 134: ...TSG 95 Service Manual 7 8...
Страница 139: ...Replaceable Mechanical Parts 8 5 TSG 95 Service Manual A1 A2 A3 Figure 1 Exploded View...
Страница 140: ...Replaceable Mechanical Parts 8 6 TSG 95 Service Manual...