Performance Verification and Adjustment Procedures
TSG 95 Service Manual
4–5
the Selected Measurements file. See your VM700A Operators manual for addi-
tional information on selecting measurements.
VM700A Auto Mode Configuration.
The following procedure sets up the VM700A
so that the Auto Measurements mode can be used.
1. Turn on the VM700A.
2. Press the following VM700A keys: Configure, <Configure Files>, and
<Measurement Locations>.
3. If you are not in the NTSC directory, press the <Switch NTSC/PAL> key.
4. Create a new NTSC Measurements Location file.
a. Press <Create File> and then <System Default>.
b. Enter “pv” for a file name then press <Done>.
NOTE
. At this point you will be in the pv file and will be able to edit it.
c.
Rotate the knob until the Auto Mode “Composite
X
XX” line is
highlighted, where XX is a number from 10 to 262.
d. Press your finger on the screen over the “XX” readout; a box should now
appear around the readout.
e.
Rotate the knob until “30” appears on the readout.
f.
Press the following keys: <Accept Input>, <Update & Exit>, and
<Leave Directory>.
5. Create a new NTSC Video_Source file.
a. Press the following keys: <Video_Source Files>, <Create File>, and
<
System Default
>.
b. Enter “pvfile” for a file name, then press <
Done
>.
c.
Rotate the knob until the “Measurements Location File: XXXXX” line
is highlighted. XXXXX will be the name of the measurements location
file currently in use.
d. Record the currently used measurements location file so that when you
have finished testing the TSG 95 you can return the VM700A to the
original settings.
e.
Press your finger over the “XXXXX” readout; a box should now appear
around the readout.
Содержание TSG 95
Страница 4: ......
Страница 6: ...TSG 95 Service Manual...
Страница 14: ...Service Safety Summary viii TSG 95 Service Manual...
Страница 29: ...Specifications TSG 95 Service Manual 1 15 Figure 1 3 PAL 75 Red and Red Field Figure 1 4 PAL 100 Red...
Страница 30: ...Specifications 1 16 TSG 95 Service Manual Figure 1 5 PAL Green Field Figure 1 6 PAL Blue Field...
Страница 31: ...Specifications TSG 95 Service Manual 1 17 Figure 1 7 PAL 50 Flat Field Figure 1 8 PAL 100 Flat Field...
Страница 34: ...Specifications 1 20 TSG 95 Service Manual Figure 1 13 PAL Modulated 5 Step Figure 1 14 PAL Multiburst...
Страница 35: ...Specifications TSG 95 Service Manual 1 21 Figure 1 15 PAL Reduced Sweep Figure 1 16 PAL Pluge...
Страница 40: ...Specifications 1 26 TSG 95 Service Manual Figure 1 25 PAL Matrix Signal UK ITS 2...
Страница 41: ...Specifications TSG 95 Service Manual 1 27 Figure 1 26 PAL Safe Area...
Страница 42: ...Specifications 1 28 TSG 95 Service Manual NTSC Waveform Diagrams Figure 1 27 SMPTE NTSC Color Bar Components...
Страница 43: ...Specifications TSG 95 Service Manual 1 29 Figure 1 28 SMPTE NTSC Color Bars Zero Setup...
Страница 45: ...Specifications TSG 95 Service Manual 1 31 Figure 1 31 NTSC Red Field Figure 1 32 NTSC Red Field Zero Setup...
Страница 46: ...Specifications 1 32 TSG 95 Service Manual Figure 1 33 NTSC Sin x x Figure 1 34 NTSC 5 Step Staircase Gray Scale...
Страница 50: ...Specifications 1 36 TSG 95 Service Manual Figure 1 41 NTSC Convergence Components...
Страница 51: ...Specifications TSG 95 Service Manual 1 37 Figure 1 42 NTC7 NTSC Composite Figure 1 43 NTC7 NTSC Combination...
Страница 52: ...Specifications 1 38 TSG 95 Service Manual Figure 1 44 FCC NTSC Composite Figure 1 45 NTSC Cable Multiburst...
Страница 53: ...Specifications TSG 95 Service Manual 1 39 Figure 1 46 NTSC Cable Sweep Figure 1 47 NTSC Matrix...
Страница 54: ...Specifications 1 40 TSG 95 Service Manual Figure 1 48 NTSC Safe Area...
Страница 104: ...Performance Verification and Adjustment Procedures 4 30 TSG 95 Service Manual...
Страница 118: ...Maintenance 5 14 TSG 95 Service Manual...
Страница 134: ...TSG 95 Service Manual 7 8...
Страница 139: ...Replaceable Mechanical Parts 8 5 TSG 95 Service Manual A1 A2 A3 Figure 1 Exploded View...
Страница 140: ...Replaceable Mechanical Parts 8 6 TSG 95 Service Manual...