Reference
Table 36.Error Code descriptions (cont.)
Rising Edge Rate / Falling Edge Rate Measurement
E901
The High level values need to
be decreased
This message is displayed if no
edge is found in the high level
Decrease the percentage of the high level
values of Rising Edge Rate/Falling Edge Rate
E902
The Low level values need to be
increased
If no edge is found in the low level
the message is displayed
Decrease the percentage of the low level
values of Rising Edge Rate/Falling Edge Rate
E903
Both the High and Low level
values need to be changed
If no edge is found in the high and
low level the message is displayed
Increase and decrease the percentage of the
high and low level values of Rising Edge
Rate/Falling Edge Rate
E904
The calculated Rising Edge
Rate is more than the calculated
bit time of the input signal
This warning appears when the
calculated Rising Edge Rate is
more than the calculated bit time of
the input signal
Increase the lower level and decrease the
upper level of the configure limits of Rising
Edge Rate/Falling Edge Rate
E905
The calculated Falling Edge
Rate is more than the calculated
bit time of the input signal
This warning appears when the
calculated Falling Edge Rate is
more than the calculated bit time of
the input signal
Increase the lower level and decrease the
upper level of the configure limits of Rising
Edge Rate/Falling Edge Rate
E906
The calculated Rising Edge
Rate and Falling Edge Rate is
more than the calculated bit time
of the input signal
This warning appears when the
calculated Rising Edge Rate and
Falling Edge Rate is more than the
calculated bit time of the input
signal
Increase the lower level and decrease the
upper level of the configure limits of Rising
Edge Rate/Falling Edge Rate
172
TDSUSB2 Universal Serial Bus Measurement Package
Содержание TDS5032
Страница 1: ...TDSUSB2 Universal Serial Bus Measurement Package 077 0015 04 www tektronix com...
Страница 12: ......
Страница 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Страница 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 88: ...Tutorial 74 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 158: ...Application Examples 144 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 225: ...Measurement Algorithms Figure 126 Equipment setup for Chirp Test TDSUSB2 Universal Serial Bus Measurement Package 211...
Страница 228: ...Measurement Algorithms 214 TDSUSB2 Universal Serial Bus Measurement Package...
Страница 239: ...Measurement Algorithms Figure 138 Viewing SOF TDSUSB2 Universal Serial Bus Measurement Package 225...
Страница 248: ...Glossary 234 TDSUSB2 Universal Serial Bus Measurement Package...