Command Descriptions
2- 312
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
WAVEAlert:SENSitivity
(TDS3000B Series and TDS3000C Series)
Sets or returns the WaveAlert anomaly detection sensitivity parameter.
Acquisition
WAVEAlert:SENSitivity { <NR1> }
WAVEAlert:SENSitivity?
<NR1>
sets the anomaly sensitivity percentage value, and ranges from 1 to 100.
WAVEALERT:SENSITIVITY?
might return
WAVEALERT:SENSITIVITY 6.0E1
, indicating that sensitivity is set
to 60%.
WAVEAlert:STATE
(TDS3000B Series and TDS3000C Series)
Sets or returns the WaveAlert testing state.
Acquisition
WAVEAlert:STATE { OFF | ON | <NR1> }
WAVEAlert:STATE?
OFF
or
<NR1>
= 0 turns off WaveAlert testing.
ON
or
<NR1>
¸
0 turns on WaveAlert testing.
WAVEAlert:STATE ON
specifies that WaveAlert waveform anomaly detection is in effect.
WAVEAlert:STATE?
returns either
0
or
1
, indicating whether WaveAlert testing is in effect.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Examples