Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 65
CH<x>:DESKew
Sets or queries the horizontal deskew time for this channel. This is equivalent to
setting
Deskew
in the
VERTICAL > Probe Setup
menu and entering a value
with the general purpose knob. Deskew lets you compensate for time delay
differences between input channels caused by signals coming in on cables of
different length.
Vertical
CH<x>:DESKew <NR3>
CH<x>:DESKew?
<x>
specifies the input channel. Valid values are 1, 2, 3, and 4.
<NR3>
is the deskew time for this channel. The range is --10.0 ns to +10.0 ns
with a resolution of 20 ps. Out of range values are clipped.
CH1:DESKEW?
might return
”-4.0E-11”
, indicating that the channel 1 deskew setting is
--40.0 ps.
CH<x>:ID? (Query Only)
Returns the probe identification including model name for level II probes.
Returns a null string for non level II probes.
Vertical
CH<x>:ID?
<x>
specifies the input channel. Valid values are 1, 2, 3, and 4.
CH1:ID?
might return
”P6243”
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Examples
Содержание TDS3000 Series
Страница 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 402: ...Appendix D Factory Initialization Settings D 4 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...