Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 203
MATH:TYPe?
ADVanced
sets the math waveform mode to advanced math. TDS3AAM only.
DPO
sets the math waveform mode to DPO math. TDS3AAM only.
DUAL
sets the math waveform mode to dual waveform math.
FFT
sets the math waveform mode to FFT math. (TDS3AAM and TDS3FFT,
available on TDS3000B Series and TDS3000C Series oscilloscopes.)
MATH:TYPE FFT
sets the math waveform mode to FFT.
MATH:DEFINE
MATH:VAR<x>
(Requires TDS3AAM)
Sets or returns the value of the user-defined math variable that can be used as
part of an advanced math waveform expression.
Math
MATH:VAR<x> <NR3>
MATH:VAR<x>?
<x>
specifies which variable to define, where <x> = 1 for VAR1 and <x> = 2 for
VAR2.
<NR3>
is a numeric value in scientific notation that sets the value of the specified
variable.
MATH:VAR1 1.251E3
sets the value of VAR1 to 1251.
MATH:VERTical:POSition
Sets or returns the math waveform display position.
Arguments
Examples
Related Commands
Group
Syntax
Arguments
Examples
Содержание TDS3000 Series
Страница 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Страница 402: ...Appendix D Factory Initialization Settings D 4 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...